CLiMET CI-3100 OPT (Ethernet) - Contact us for pricing

Zoom

CLiMET CI-3100 OPT (Ethernet)

The new CLiMET CI-3100 OPT (Ethernet) particle counter is the latest instrument offered by OptiCal Sciences that includes Ethernet connectivity. The instrument can be fully controlled by your choice of SCADA software and is compatible with industry standard TCP/IP Modbus communication protocols. This particle counter comes with an internal vacuum pump and continuously counts particles from a fixed position on the network. CI-3100 is capable of measuring the essential two channels, 0.5 and 5 microns at 1 cubic foot per minute and comes in pharmaceutical grade stainless steel enclosure.

 

Applications

  • Continuous monitoring in pharmaceutical manufacturing and hospital clean room environments to meet EU GMP annex 1 requirements.
  • Point of fill monitoring in pharmaceutical manufacturing.
  • To meet monitoring requirements in clean room environments.

Benefits

  • Ethernet Modbus OPT allows easy integration with your choice of data acquisition platform.
  • Built in vacuum pump eliminates the need for an external vacuum source.
  • Measures the essential two channels - 0.5 and 5 microns.
  • Low particle emission ensuring that the counter does not contaminate your environment.
  • Emission certificate for particle emissions from the counter issued with every counter.
  • Sample volume settings of either 1 cubic foot or 1 cubic meter with nominal sample times of 60 or 2118.88 seconds respectively.
  • Latch or Rolling count option enabling quicker results.
  • Pharmaceutical grade stainless steel enclosure.
  • Ideally suited for Grade A & B continuous monitoring requirements to meet EU GPM annex 1 requirements.
  • Built-in website with individual web addresses enabling easy viewing of operating data using a web browser.

 

Downloads

Downloadable CI-3100 Ethernet OPT Brochure
(To save the file right click on the download link, then select "Save Target As...".)

 

Additional Information

  • Our service engineers are fully qualified to calibrate the CI-3100 OPT to meet national standards including the latest ISO 21501-4. Please refer to our Calibration and Repairs section for more information.
  • Service contracts available to cover routine calibration visits, emergency callouts and replacement parts. Please refer to our Service Contracts section for more information.
  • OptiCal Sciences can provide Particle Monitoring Systems to continuously sample your clean environment during sterile manufacturing. Please refer to our Particle Monitoring Systems section for more information.
  • We offer validation service for Particle Monitoring Systems. Please refer to our Validation section for more information.

 

Frequently Asked Questions (FAQ)

What is the size sensitivity of the instrument?
The CI-3100 OPT (Ethernet) model has size sensitivity of 0.5µm and 5.0µm.

What is the flow rate / sample volume of the instrument?
The CI-3100 OPT (Ethernet) model has a flow rate of 1.0 Cubic Feet per Minute with automatic flow control set at the factory. The 1 cubic foot or 1 cubic meter, nominal sample times are 60 seconds or 2118.88 seconds but may vary based on flow and laser status during the sample. Sample data is updated once per sample volume continuously; rolling data is updated 60 times per volume (nominally every second or 35.3 seconds).

What can be used to clean the instrument?
IPA or hydrogen peroxide can be used as long as the pump is turned off and the particle inlet is capped.

How long is the instrument covered for under warranty?
The CI-3100 OPT (Ethernet) model comes with a 1 year warranty covered by the manufacturer.

Contact us for pricing